بازخورد درباره این کالا

دتکتور FLD مدل 474 دستگاه HPLC ساخت واترز

0
از 0 رای

0دیدگاه

پرتوژن
ضمانت اصالت کالا

تماس بگیرید

مرا اگاه کن


از طریق:
ثبت

نقد و بررسی

دتکتور FLD مدل 474 دستگاه HPLC ساخت واترز

دتکتور FLD مدل 474

Condition Specification
Monochromator Holographic concave diffraction grating (for both
excitation and emission)
Light source 150 W Xenon lamp
Wavelength range 200 to 700 nm (for excitation and emission). Can be set from 0 to 900
Spectrum band width Excitation: 18 nm
Emission: 10, 18, or 40 nm (set on keypad)
Wavelength accuracy ±2 nm
Wavelength repeatability ±0.3 nm
Detector Excitation: Photodiode
Emission: Photomultiplier tube
Second order filter Automatically inserted into light path when Ex > 400 Nm
Flow cell 16 µL (standard) or 5 µL
Maximum flow cell pressure 16 µL: 145 psi (10 bar)
5 µL: 500 psi (34.5 bar)
Wetted surfaces Synthetic silica, fluororesin, PEEK, stainless steel
SXS-316
Sensitivity S/N for the Raman peak of water is 320 or above,
where Ex λ = 350 nm and time constant is 1.5 sec,
16 µL cell (Instruments shipped prior to June 1997,
S/N for the Raman peak of water is 200 or above)
Attenuation range (Recorder output) Ten steps, including 1, 2, 4, 8, 16, 32, 64, 128, 256, S
Gain range 1, 10, 100, 1000
Response Fast, Standard, Slow, and digital filter
Signal processing Digital processing by A/D and D/A converters (with
surrounding temperature compensation circuits)
Outputs • Recorder: 10 mV full-scale (can be inverted)
• Integrator: 1 V full-scale
• Marker output
• Leak output
Inputs • Marker input
• Auto Zero input
• Program reset
Program functions Time program with following variables:
• Wavelength change (excitation and emission)
• Gain
• Attenuation
• Wavelength scan
• Ex, Em Spectra
Wavelength scan functions Manual and time program parameters:
• Time
• Interval
• Amplitude
• Memory location
• Ex, Em Spectra
Spectrum storage Ten locations for both emission and excitation spectra
Self-test diagnostics • Memory (ROM and RAM)
• DC power
• Excitation energy decrease
• Leak in cell
• Lamp use time
Lamp use timer Internal timer for accumulating the total hours on the Xenon lamp
Ambient temperature 10 °C to 32 °C (50 °F to 90 °F) operation
10 °C to 60 °C (50 °F to 140 °F) storage
Relative humidity 20 to 80%, non-condensing

 

جهت مشاوره و خرید انواع تجهیزات آزمایشگاهی با مشاوران واحد فروش شرکت شناسا پرتوژن ۷۵۳۱۱ – ۰۲۱ در ارتباط باشید.

امتیاز کاربران به:

دتکتور FLD مدل 474 دستگاه HPLC ساخت واترز (0نفر)
کیفیت ساخت
0
ارزش خرید به نسبت قیمت
0
امکانات و قابلیت ها
0
سهولت استفاده
0
دیدگاه خود را در باره این کالا بیان کنید برای ثبت نظر، لازم است ابتدا وارد حساب کاربری خود شوید. اگر این محصول را قبلا از این فروشگاه خریده باشید، نظر شما به عنوان مالک محصول ثبت خواهد شد. افزودن دیدگاه
نظرات کاربران

هیچ دیدگاهی برای این محصول نوشته نشده است.

    هیچ پرسش و پاسخی ثبت نشده است.

پرسش خود را درباره این کالا بیان کنید

ثبت پرسش

محصولات مرتبط